sales@gninstruments.com

                                                      Type:CX-LSM-1609A laboratory sodium meter

                                                      Illustrate:It can be widely used for continuous monitoring of sodium ions of the solutions in thermal power plants, chemical fertilizer, metallurgy, environmental protection, pharmaceutical industry, biochemical industry, foodstuff and running water.

                                                      CX-LSM-1609A Laboratory Sodium Meter


                                                      Brief Introduction
                                                      1.CX-LSM-1609Ais made based on the improvement of the previous pointer scale display, knobs, switches and non-intelligence design.
                                                      2.It is equipped with the dedicated laboratory PPB sodium ion measurement electrode and the reference electrode, with the static measurement performance in the beaker greatly improved, compared with the traditional sodium meters.
                                                       
                                                      Main Features
                                                      1.Double-row nixie tube: To display μg/L, pNa/mV and temperature simultaneously.
                                                      2.Computer-based: To adopt high-performance CPU chip, high-precision AD conversion technology and 8usMTtechnology to complete multi-parameter measurement and temperature compensation, with high precision and good repeatability
                                                      3.High reliability: Single board structure, touch button, no switch knob
                                                      4.Input impedance :> 1012Ω
                                                      5.High sensitivity: In the static water sample, it can detect 0.2ppb (8pNa)
                                                      6.Lasting stability: There is a very stable reading.
                                                      7.Waterproof and anti-erosion design
                                                      8.Network function: RS232 bi-directional communication, equipped with a printer or TLab8.0 communications and
                                                      data processing software.

                                                      Main Parameters

                                                      Display Double-row nixie tubes display
                                                      Measuring range 0.2 ug/L~23000mg/L,0.00 Pna~8.00 PNa.-1000~ +1000 mV. 0~99.9°C
                                                      Resolution 0.1 ug/L; 0 .1 mV, 0.1°C
                                                      Repeatability reading ± 2.5%
                                                      Accuracy  reading ± 2.5%, ±0 .2 mV, ±0.3°C
                                                      Continuous stability time >100 seconds
                                                      Detection lower limit 0 .2 ug/L (8pNa)
                                                      Input impedance >1012Ω
                                                      Temperature compensation range automatic/manual 0 ~ 99.9 °C
                                                      Response speed 32 seconds (25°C) or 50 seconds (15°C)
                                                      Communication interface RS232 bi-directional communication
                                                      Power supply AC 220V ± 22V, 50Hz ± 1Hz, external power supply
                                                      Power-down memory > 5 years
                                                      Calibration mode One-point calibration, two-point calibration
                                                      Working conditions ambient temperature: 0-60°C;  relative humidity <≤90
                                                      Standard configuration:One secondary instrument; one power supply; one reference electrode, one measuring electrode, one temperature electrode; a set of electrode support, a set of standard buffer solution