sales@gninstruments.com

                                                      Type:CX-ICM-1101 industrial conductivity meter

                                                      Illustrate:Widely used for continuous monitoring of the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy, environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.

                                                      CX-ICM-1101 industrial conductivity meter

                                                      Brief Introduction
                                                      1.It is our company’s latest top-class instrument equipped with a microprocessor
                                                      2.It is characterized by its intelligent operation, multi-function and high measuring performance and its strong environmental adaptability
                                                       
                                                      Main Features
                                                      1.Display:conductivity, output current, temperature, time and status.
                                                      2.Bitmap type liquid crystal display modulewith high resolution is adopted.
                                                      3.Thereis no symbol or code that is defined by the manufacturer.
                                                       
                                                      Main Parameters

                                                      Measuring range 0.01~20μS/cm(Electrode:K=0.01),00.1~200μS/cm
                                                      (Electrode: K=0.1)
                                                      1.0~2000μS/cm(Electrode:K=1.0),10~20000μS/cm
                                                      (Electrode:K=10.0)                       
                                                      30~600.0mS/cm  (Electrode: K=30.0)
                                                      Intrinsic error of the electronic unit 2. : conductivity: ±0.5%FS, temperature: ±0.30C
                                                      Automatic temperature compensation 0~99.9C, with 25C as the reference temperature
                                                      Water sample tested 0~99.9C, 0.6MPa
                                                      Intrinsic error of the instrument Conductivity: ±1.0%FS, temperature: ±0.5C
                                                      Automatic temperature compensation error ±0.5%FS
                                                      Repeatability error ±0.2%FS±1 Unit
                                                      Stability ±0.2%FS±1 unit/24h
                                                      Isolated current output 0~10mA ( load<1.5kΩ),4~20mA (load<750Ω)
                                                      Output current error ≤±l%FS
                                                      Error caused by ambient temperature ≤±0.5%FS
                                                      Error caused by the supply voltage  ≤±0.3%FS
                                                      Alarm relay  AC 220V, 3A
                                                      Communication interface RS485 or 232 (optional)
                                                      Power supply AC 220V±22V, 50Hz±1Hz
                                                      Protection grade IP65
                                                      Clock accuracy ±1 minute/month
                                                      Data storage capacity l month (1 point/5 minutes)
                                                      The time for Saving data 10 years
                                                      Overall dimension 146 (length) x 146 (width) x 108 (depth) mm;
                                                      Dimension of the hole 138 x 138mm
                                                      Working conditions ambient temperature0~60C; relative humidity <85%
                                                      Note: The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, and 30.0.