sales@gninstruments.com

                                                      Type:CX-ICM-1101A industrial conductivity meter

                                                      Illustrate:It is applicable for pure and ultrapure water treatment, industrial and sewage water treatment in the following industries: electric power, environmental protection, chemical industry, petroleum, paper making, metallurgy, semi-conductor, medicine, foodstuff, etc

                                                      CX-ICM-1101A industrial conductivity meter

                                                      Brief Introduction
                                                      1.It is a kind of on-line monitoring instrument for water quality analysis equipped with a microprocessor.
                                                      2.Widely used for continuous monitoring of the conductivities of demineralized water, raw water, steam condensate, distilled sea water and de-ionized water.
                                                       
                                                      Main Features
                                                      1.Large vacuum fluorescent display (VFD), and man-machine interaction function.
                                                      2.The functions of automatic temperature compensation and automatic range-switching.
                                                      3.Available with history curve, real-time trend chart and clock.
                                                      4.There are four kinds of built-in recorders. It has the functions of isolated current output and alarm.
                                                      5.The user can set the password or consult the “Help” menu or the “Service Guide” Menu.
                                                      6.It has power-off protection. The data can be saved for 5 years under a continuous power-failure condition.
                                                      7.The communication interface can be connected directly to a computer and enables remote operation and control.
                                                      8.The instrument adopts universal module power supply.
                                                      9.It is equipped with rib-type, plug-in and push-in types of conductivity electrode with a electrode constant of
                                                      K=0.01, K=0.1 or K=1.0.
                                                      10.It adopts titanium electrode that can withstand oxidization and erosion and can be easily used and maintained.
                                                       
                                                      Main Parameters

                                                      Measuring range 0~ 20 μS.cm-1 ,0~200 μS.cm-1, 0~2000μS.cm-1, 0~20mS.cm-1
                                                      Automatic temperature compensation 0-60℃ (with 25℃ as the reference temperature)
                                                      Intrinsic error Conductivity:±0.5%F.S±1 unit
                                                      Error of the temperature compensation ±0.5%F.S
                                                      Stability ±0.2/F.S±1 unit/24h
                                                      Repeatability error ≤0.2%F.S±1 unit
                                                      Output current error ±1%F.S
                                                      Alarm error of the electronic units ±2%F.S
                                                      Water sample pressure 0.6MPa
                                                      Intrinsic error of the instrument: conductivity ±1%F.S, temperature: ±0.5℃
                                                      Maximum output load and error For 0~10mA, the maximum output load is l.5K Ω±1%F.S.
                                                      For 4~20mA, the maximum output load is 750Ω±1%F.S.
                                                      Power supply 85~265VAC 50±1Hz
                                                      Overall dimension 151 (length) x 151 (width) x 251 (height) mm
                                                      Dimension of the mounting hole 138 x 138mm
                                                      Connection pattern of inlet/ outlet water pipes Φ6,Φ8,Φl0,Φl2 and Φ14.
                                                      Weight 1.2kg
                                                      sensor types and installation methods Flow-through type,duct-type,immersed type